Duplex PNP50 Two-Stage Gas Manifold

Duplex PNP50 Two-Stage Gas Manifold

Manifolds and stabilizersGCE Healthcare

The GCE Duplex PNP50 is a two-stage gas manifold, with an integrated third source entry point for LOX, compressor, or 3rd backup source.

USE OUR CONFIGURATOR

Duplex PNP50 Two-Stage Gas ManifoldIntroduction

The GCE Duplex PNP50 two-stage manifold is engineered for hospital pipeline systems, offering a reliable medical gas source with an integrated third source entry point for LOX, compressors, or other backup sources. This manifold reduces gas pressure in two steps to ensure a constant distribution pressure. It features an innovative touchscreen alarm displaying all three sources. The Duplex PNP50 gas manifold supports medical gases from cylinders of 300 bar, 200 bar, or 150 bar, providing a fully automatic changeover system with advanced communication protocols for seamless integration.

Duplex PNP50 Two-Stage Gas ManifoldFeatures

Comprehensive Management

  • One panel for managing all three gas sources
  • Suitable for all medical gases

Versatile Compatibility

  • Available for 300-bar, 200-bar, or 150-bar cylinders
  • LOX, compressor, or third cylinder source connection port

Automated Functionality

  • Fully automatic changeover system
  • Visual indication in alarm for an active source
  • New pneumatic side switchover

Advanced Monitoring

  • Integrated communication protocols MODBUS TCP/IP and RS485 RTU
  • All sources monitored by the integrated alarm
  • Monitoring leakage on the non-active source (leaking reserve side)

Enhanced Control

  • Touchscreen alarm
  • Optional position control on the outlet shut-off valve
  • Optional terminal unit test point

Maintenance and Installation

  • Service and maintenance can be performed during manifold operation (no gas supply interruption)
  • Easy installation due to preinstalled backplate

Duplex PNP50 Two-Stage Gas ManifoldGeneral properties

General properties

Gas type(s)O2, Air, Air-800, N2O, CO2
Nominal flow (m3/h)50
Cylinder inlet pressure300 bar (30 000 kPa) / 200 bar (20 000 kPa)
LOX or Compressor inlet pressure8 - 14 bar (800 - 1 400 kPa)
3rd source standby manifold pressure7 - 10 bar (700 - 1 000 kPa)
Outlet nominal pressure4.5 bar ( 450 kPa)
Inlet connection2x W21,8×1/14"M, 1xM20x1F
Outlet connectionG3/4“M
Outlet pressure relieve valve6.8 bar (including manual activation)
Intermediate pressure relieve valve17 bar (including manual activation)
Pressure relieve valve pipe dimensionø 15 mm
Testing pointNIST QC or QC in national standard
Alarmintegrated with touch screen
StandardMDD 93/42/EEC, EN ISO 7396-1, EN 60601-1-2, SIS HB 370

Quality Standards

We comply with all relevant regulations

See our CERTIFICATIONS >

Duplex PNP50 Two-Stage Gas ManifoldDownloads