The manifold Duplex SB50 is designed to be used as a third source of supply in medical central gas systems. The manifold will deliver gas when the nominal supply system pressure falls below a set level (7 bar). This is a back up source.
Together with Duplex SB50, we recommend to use the Duplex PNP50 unit
and alarm providing all alarms according to standard.
FOR A COMPLETE INSTALLATION ADD:
> Collecting pipe set (high pressure valves, and non-return valves / high pressure components)
> Cylinder retaining brackets (included in gas cylinder collecting pipe set)
> High pressure hoses with safety wire
> Plug for close collecting pipeline
> Gas name sign
Together with Duplex SB50, we recommend to use the Duplex PNP50 unit
and alarm providing all alarms according to standard.
FOR A COMPLETE INSTALLATION ADD:
> Collecting pipe set (high pressure valves, and non-return valves / high pressure components)
> Cylinder retaining brackets (included in gas cylinder collecting pipe set)
> High pressure hoses with safety wire
> Plug for close collecting pipeline
> Gas name sign
Automatic Backup:
Seamlessly activates to maintain system pressure in emergency situations.
Compatibility:
Designed for optimal performance with the Duplex PNP50 unit and alarm, providing all necessary alerts per regulatory standards.
Complete Installation Options:
Collecting pipe set with high-pressure valves and non-return components
Cylinder retaining brackets
High-pressure hoses with safety wire
Gas name signs
Automatic Backup:
Seamlessly activates to maintain system pressure in emergency situations.
Compatibility:
Designed for optimal performance with the Duplex PNP50 unit and alarm, providing all necessary alerts per regulatory standards.
Complete Installation Options:
Collecting pipe set with high-pressure valves and non-return components
Cylinder retaining brackets
High-pressure hoses with safety wire
Gas name signs
Gases: | O2, Air, N2, N2O, CO2 (all medical gases) |
---|---|
Nominal flow: | 50 m3/h |
Outlet nominal pressure: | 7 bar (setting range 7–15 bar) |
Inlet connection: | W21,8×1/14“M |
Outlet connection: | G3/4“M + soldering piece pipe ø 15 mm |
Pressure relieve valve: | 17 bar including manual activation |
Pressure relive valve pipe dimension: | ø 10 mm |
Purge valves connection: | W21,8×1/14“M + soldering piece pipe ø 10 mm |
Regulatory status: | MDD 93/42/EEC, EN ISO 7396-1, EN 60601-1-2, present SIS HB 370 |
Loading... ... loading table ... |
Item No. | Denomination | Gas |
---|---|---|
073430001 | DUPLEX SB50 O2 7B GCE 000 | O2 |
073430008 | DUPLEX SB50 AIR GCE | Air |
073430009 | DUPLEX SB50 N2O GCE | N2O |
073430010 | DUPLEX SB50 CO2 GCE | CO2 |