The GCE ECOlite® 4000 electronic oxygen conserving device enables patient-friendly and efficient oxygen therapy treatment.
The GCE ECOlite® 4000 electronic oxygen conserver is designed to improve patient care and optimize resource usage. Delivering oxygen only during the inspiration phase of the breathing cycle allows for savings of up to 10 times compared to traditional continuous-flow oxygen therapy.
The ECOlite® 4000 features both automatic and manual flow rate settings, providing flexibility to meet diverse patient needs. This oxygen conserving device is equipped with visual and audio alarms, ensuring immediate notification of any issues for enhanced safety.
The GCE ECOlite® 4000 electronic oxygen conserver is designed to improve patient care and optimize resource usage. Delivering oxygen only during the inspiration phase of the breathing cycle allows for savings of up to 10 times compared to traditional continuous-flow oxygen therapy.
The ECOlite® 4000 features both automatic and manual flow rate settings, providing flexibility to meet diverse patient needs. This oxygen conserving device is equipped with visual and audio alarms, ensuring immediate notification of any issues for enhanced safety.
Advanced Technology
The ECOlite® 4000 electronic oxygen conserver uses advanced technology to precisely deliver oxygen exclusively during the inspiration phase of breathing.
Automatic And Manual Flow Rates
This conserving device for oxygen offers both automatic and manual flow rate settings for flexibility and accurate control.
Fixed Flow Rate
The fixed flow rate ensures consistent oxygen delivery.
Visual and Audio Alarms
The ECOlite® 4000 digital oxygen conserver notifies via visual and audio alarms in case of issues.
Durable Construction
Designed to withstand the demands of clinical environments, ensuring long-lasting performance.
Environmental Protection
A reliable and sustainable choice.
Advanced Technology
The ECOlite® 4000 electronic oxygen conserver uses advanced technology to precisely deliver oxygen exclusively during the inspiration phase of breathing.
Automatic And Manual Flow Rates
This conserving device for oxygen offers both automatic and manual flow rate settings for flexibility and accurate control.
Fixed Flow Rate
The fixed flow rate ensures consistent oxygen delivery.
Visual and Audio Alarms
The ECOlite® 4000 digital oxygen conserver notifies via visual and audio alarms in case of issues.
Durable Construction
Designed to withstand the demands of clinical environments, ensuring long-lasting performance.
Environmental Protection
A reliable and sustainable choice.
Settings | Manual/Automatic |
---|---|
Triggering | At each breath |
Sensitivity | 0.13 cm H2O |
Regulating Pressure | 1.6 Bar |
Accuracy | 0.5–1.5 l/min +/- 30% / 2–8 l/min +/- 15% |
Cycle Output | 0.5 to 8 l/min corresponding to 5–80 ml per bolus |
Alarms | Low battery / No oxygen supply / No inhalation |
Battery | RO6, AA, Alkaline 1.5 V |
Oxygen Supply Pressure | Between 1.6 and 5 Bar |
Flow | Minimum 4 liters per minute |
Dimensions | H-101 mm / W-85mm / D-32mm |
Weight | 184g / 0.4lbs without battery |
Operational | - 10°C to +40°C (-50°F to +104°F) |
Storage | - 40°C to +70°C (-104°F to +158°F) |
Relative Humidity | 25% to 95% |
Loading... ... loading table ... |
Item No. | Description |
---|---|
325197478 | ECOlite® 4000 COMPL DE |
325197479 | ECOlite® 4000 COMPL UK |
325197544 | ECOlite® 4000 COMPL SE |
325197545 | ECOlite® 4000 COMPL FRA |
325197617 | ECOlite® 4000 COMPL 1.6 BAR |
Code | Description |
---|---|
14111220 | Standard Nasal cannula |
14111222 | Spiral Hose SE |
14090329 | Spiral Hose UK |
14090417 | Spiral Hose 1.6 bar |
14090510 | Spiral Hose DE |
325197699 | Supply hose ECOlite® 4000 3/8 |
14090535 | ECOlite® 4000 Carry Bag |
14090631 | ECOlite® 4000 Carry Bag Trolley |
325112719 | Belt bag |
325112670 | Battery cover |