The gas cylinder manifold MC25 primarily intended for small and medium-sized hospitals.
The gas cylinder pressure is regulated in two steps. The change-over between operating side and reserve side is made automatically without any diff erences in the operating pressure.
The alarm signal comes from the pressure switches to the alarm unit. The alarm signals from the alarm unit can be forwarded directly to a monitoring desk. Function control and service can be carried out without interruption in the gas supply.
The gas cylinder pressure is regulated in two steps. The change-over between operating side and reserve side is made automatically without any diff erences in the operating pressure.
The alarm signal comes from the pressure switches to the alarm unit. The alarm signals from the alarm unit can be forwarded directly to a monitoring desk. Function control and service can be carried out without interruption in the gas supply.
MC25 includes the following components:
MC25 includes the following components:
Gases | O2, Air, N2, N2O, CO2 |
---|---|
Nominal flow (m3/h) | 25 |
Inlet nominal pressure | 200 (20 000 kPa) |
Outlet nominal pressure | 4,5 bar (setting range 0,5–6 bar) |
Inlet connection | W21,8×1/14"M |
Outlet connection | G1/2"M + soldering piece pipe ø 10, ø 15 mm |
Outlet pressure relieve valve | 6,8 bar |
Intermediate pressure relieve valve | 17 bar |
Pressure relieve valve pipe dimension | ø 15 mm |
Loading... ... loading table ... |
Item No. | Denomination | Gas | PRV | Signalization |
---|---|---|---|---|
0732300 | MC25 – 2×1 | O2 | With pull-ring | Alarm TMA6703 |
0732301 | MC25 – 2×1 | Air | With pull-ring | Alarm TMA6703 |
0732302 | MC25 – 2×1 | N2O, CO2 | With pull-ring | Alarm TMA6703 |