Capacity of 90 m3/h
The MM90 HP unit medical manifold is intended for use in hospital pipeline system as medical gas source. Together with MM90, always use an alarm providing all alarms according to standard (like gas alarm). As 2nd stage is recommended to use a stabilizer. The manifold will deliver gas from the operating bank to the manifold pressure regulator until the cylinders are exhausted. At that point the supply will switch to the reserve bank and the empty bank can be replenished. The object gives uninterrupted gas supply. Gas Alarm is a standard accessory. The gas alarm gives visual and audible indication.
The MM90 HP unit medical manifold is intended for use in hospital pipeline system as medical gas source. Together with MM90, always use an alarm providing all alarms according to standard (like gas alarm). As 2nd stage is recommended to use a stabilizer. The manifold will deliver gas from the operating bank to the manifold pressure regulator until the cylinders are exhausted. At that point the supply will switch to the reserve bank and the empty bank can be replenished. The object gives uninterrupted gas supply. Gas Alarm is a standard accessory. The gas alarm gives visual and audible indication.
It surveils and the alarm sounds when the following happens:
The Gas alarm is able to communicate with other equipment through relays.
MM90 INCLUDES THE FOLLOWING COMPONENTS:
FOR A COMPLETE MM90 HP UNIT MANIFOLD ADD:
Complies with Medical Devices Directive 93/42/EEC
Complies with EN ISO 7396-1 (Central Gas Supply Systems)
Complies with EN 60601-1-2 (Electromagnetic compatibility)
present SIS HB 370 and HTM 02-01
It surveils and the alarm sounds when the following happens:
The Gas alarm is able to communicate with other equipment through relays.
MM90 INCLUDES THE FOLLOWING COMPONENTS:
FOR A COMPLETE MM90 HP UNIT MANIFOLD ADD:
Complies with Medical Devices Directive 93/42/EEC
Complies with EN ISO 7396-1 (Central Gas Supply Systems)
Complies with EN 60601-1-2 (Electromagnetic compatibility)
present SIS HB 370 and HTM 02-01
Gas type(s) | O2, Air, N2, N2O, CO2 |
---|---|
Nominal flow (m3/h) | 90 |
Inlet nominal pressure | 200 (20 000 kPa) |
Outlet nominal pressure | 9 bar (setting range 9 – 15 bar) |
Inlet connection | W21,8×1/14"M |
Outlet connection | G1/2"M + soldering piece pipe ø 22 mm |
Intermediate pressure relieve valve | 16 bar |
Pressure relieve valve pipe dimension | ø 10 mm |
Purge valves connection | W21,8×1/14"M + soldering piece pipe ø 10 mm |
Loading... ... loading table ... |
Item No. | Description | Gas | PRV | Signalization |
---|---|---|---|---|
0732322 | MM90 – HP unit AUTO 2×1 | O2 | Standard | Alarm TMA6703 |
0732323 | MM90 – HP unit AUTO 2×1 | Air | Standard | Alarm TMA6703 |
0732324 | MM90 – HP unit AUTO 2×1 | N2O, CO2 | Standard | Alarm TMA6703 |
0727308* | MM90 – HP unit AUTO 2×1 | O2, N2O, Air, CO2, N2 | Standard | *basic version without electric sensors |
0727309 | MM90 – HP unit AUTO 2×1 | O2 | Standard | Pressure switch |
0727310 | MM90 – HP unit AUTO 2×1 | Air | Standard | Pressure switch |
0727311 | MM90 – HP unit AUTO 2×1 | N2O, CO2 | Standard | Pressure switch |
You can have confidence in our global network
Copyright © 2024 GCE. All rights reserved.
GCE Group
Turmstrasse 30
6321 Steinhausen
Zug
Switzerland