The Simplex MMR gas manifold is suitable for such health care where the capacity requirement is limited, such as laboratories and small health care clinics, veterinary etc
This gas manifold consists of only one group of cylinders.
The regulator is mounted in the collection unit. Each inlet connection has a fi lter, a non-return valve and a shut-off valve. This arrangement makes it possible to use one cylinder at a time.
In order to obtain a stabile outlet pressure this gas manifold is equipped with a preset two-stage regulator. On the high pressure side of the regulator there is a contact gauge the signal of which can be carried further to an alarm unit.
This gas manifold consists of only one group of cylinders.
The regulator is mounted in the collection unit. Each inlet connection has a fi lter, a non-return valve and a shut-off valve. This arrangement makes it possible to use one cylinder at a time.
In order to obtain a stabile outlet pressure this gas manifold is equipped with a preset two-stage regulator. On the high pressure side of the regulator there is a contact gauge the signal of which can be carried further to an alarm unit.
SIMPLEX MMR INCLUDES THE FOLLOWING COMPONENTS:
FOR A COMPLETE SIMPLEX MMR ADD:
SIMPLEX MMR INCLUDES THE FOLLOWING COMPONENTS:
FOR A COMPLETE SIMPLEX MMR ADD:
Gas type(s) | O2, Air, N2, N2O, CO2 |
---|---|
Nominal flow (m3/h) | 30 |
Inlet nominal pressure (BAR) | 200 (20000 kPa) |
Outlet nominal pressure | 5 bar (setting range 4–5 bar) |
Inlet connection | W21,8×1/14"M |
Outlet connection | G3/8"M |
Pressure relieve valve | 6 bar |
Pressure relieve valve pipe dimension | ø 8 mm |
Purge valves connection | W21,8×1/14"M+pipe ø 15 mm |
Loading... ... loading table ... |
Item No. | Denomination | Gas | RPV | Signalization |
---|---|---|---|---|
325397702 | Simplex MMR | O2, Air, N2, Ar, N2O, CO2 | Standard integrated | Contact gauge |