MM40 - HP Unit

MM40 - HP Unit

Manifolds and stabilizersGCE Healthcare

Manifold MM40 HP unit is an automatic manifold

USE OUR CONFIGURATOR

MM40 - HP UnitIntroduction

It is working on the principal of different pressures between the operation and reserve regulator. By the manual lever, the operator can decide which side will be the operational side and which will be the reserve side. When the operating side is empty, the manifold will without any action start to supply gas from the reserve side with the lower regulator pressure and fulfill the requests to supply without interrupting the flow.

MM40 - HP UnitFeatures

  • Gas Alarm is a standard accessory
  • The gas alarm gives visual and audible indications
  • The gas alarm is able to communicate with other equipment through relays
  • Manifold MM40 is only first stage of regulation and must be installed together with a Stabilizer which will stabilize the final pressure used in the hospital gas outlets
  • GCE medical manifolds are CE-marked and fulfill the ISO 7396-1 standard

MM40 - HP UnitGeneral properties

General properties

GasesO2, Air, N2, N2O, CO2
Nominal flow (m3/h) 40
Inlet nominal pressure (BAR)200 (20000 kPa)
Outlet nominal pressure 12 bar (setting range 9–16 bar)
Inlet connectionW21,8×1/14"M
Outlet connectionG1/2"M + soldering piece pipe ø 10, ø 15 mm
Pressure relieve valve17 bar
Pressure relieve valve pipe dimension ø 10 mm
Purge valves connectionW21,8×1/14"M

Quality Standards

We comply with all relevant regulations

See our CERTIFICATIONS >

MM40 - HP UnitDownloads

Find our Products

Source our products wherever you are

You can have confidence in our global network

Find a distributor
GCE main LogoPart of ESAB Corporation

GCE Group

Turmstrasse 30

6321 Steinhausen

Zug

Switzerland

Copyright © 2024 GCE. All rights reserved.
facebooktwitterlinkedinyoutubeinstagram