MC80 - HP UNIT + MC80 - STABILIZER + GAS ALARM SYSTEM
1. MC80 - HP UNIT
This module contains two regulators with pressure relieve valves and it is connected to two various cylinder banks with high pressure hoses. When the cylinder bank, which has been connected for operation, has been emptied the other duty side is automatically connected.
2. MC80 - STABILIZER
The stabilizer makes the operating pressure in the distribution system remain constant. The module contains two regulators with pressure relieve valves. Since the gas pressure is reduced in two steps the drop in pressure, when changing from the operating cylinder to the other bank of cylinders, is kept to a minimum. The unit is prepared for connection to a liquid oxygen supply tank (LOX).
3. GAS ALARM SYSTEM - based on the product variant
Alarm systems from GCE are user friendly solutions, with simple control and lots of extra functionality. It surveils electronically and the alarm sounds when the following happens:
1. Too high or too low distributing pressure,
2. Too high intermediate pressure,
3. Leakage on the reserve gas cylinder bank,
4. When change of operating side has been effected.
When connected to a liquid tank the following disturbances will be reported:
1. Too high or too low distribution pressure,
2. Too high intermediate pressure,
3. Leakage from the reserves,
4. When change of operating side has been effected.
The product is either equipped with an alarm system or a sensor only.
1. MC80 - HP UNIT
This module contains two regulators with pressure relieve valves and it is connected to two various cylinder banks with high pressure hoses. When the cylinder bank, which has been connected for operation, has been emptied the other duty side is automatically connected.
2. MC80 - STABILIZER
The stabilizer makes the operating pressure in the distribution system remain constant. The module contains two regulators with pressure relieve valves. Since the gas pressure is reduced in two steps the drop in pressure, when changing from the operating cylinder to the other bank of cylinders, is kept to a minimum. The unit is prepared for connection to a liquid oxygen supply tank (LOX).
3. GAS ALARM SYSTEM - based on the product variant
Alarm systems from GCE are user friendly solutions, with simple control and lots of extra functionality. It surveils electronically and the alarm sounds when the following happens:
1. Too high or too low distributing pressure,
2. Too high intermediate pressure,
3. Leakage on the reserve gas cylinder bank,
4. When change of operating side has been effected.
When connected to a liquid tank the following disturbances will be reported:
1. Too high or too low distribution pressure,
2. Too high intermediate pressure,
3. Leakage from the reserves,
4. When change of operating side has been effected.
The product is either equipped with an alarm system or a sensor only.
DUPLEX (MC80) INCLUDES THE FOLLOWING COMPONENTS:
FOR A COMPLETE DUPLEX MC80 STANDBY MANIFOLD ADD:
DUPLEX (MC80) INCLUDES THE FOLLOWING COMPONENTS:
FOR A COMPLETE DUPLEX MC80 STANDBY MANIFOLD ADD:
Gases | O2, Air, N2, N2O, CO2 |
---|---|
Nominal flow (m3/h) | 200 |
TECHNICAL DATA - HIGH PRESSURE UNIT MC80 | |
Inlet nominal pressure (BAR) | 200 (20 000 kPa) |
Outlet nominal pressure | 12 bar (setting range 10–16 bar) |
Inlet connection | W21.8×1/14"M |
Outlet connection | G3/4"F |
Pressure relieve valve | 17 bar |
Pressure relieve valve pipe dimension | ∅ 10 mm |
Purge valves connection | W21.8×1/14"M + pipe ∅ 15 mm |
TECHNICAL DATA - STABILIZER MC80 | |
Inlet maximal pressure | 20 bar (2000 kPa) |
Outlet nominal pressure | 4,5 bar (setting range 0,5–6 bar) |
Inlet connection | G3/4"F |
Outlet connection | G3/4"F |
Pressure relieve valve | 6,8 bar |
Pressure relieve valve pipe dimension | ø 15 mm |
Loading... ... loading table ... |
Item No. | Denomination | Gas | RPV | Signalization |
---|---|---|---|---|
0732328 | DUPLEX 2×1 | O2 | Manual activation | Alarm TMA9705 |
0732329 | DUPLEX 2×1 | Air | Manual activation | Alarm TMA9705 |
0732330 | DUPLEX 2×1 | N2O/CO2 | Manual activation | Alarm TMA9705 |
You can have confidence in our global network
Copyright © 2024 GCE. All rights reserved.
GCE Group
Turmstrasse 30
6321 Steinhausen
Zug