Manifold MM40 – STABILIZER is a second stage pressure reduction unit with the task to equalize the eventual pressure variation in the hospital pipeline system to ensure a correct pressure from the terminal units.
MM40 – STABILIZER is only a second stage reduction unit where the primary gas supply is provided by high pressure gas manifolds (such as MM40 – HP Unit). In case of a signal for pressure deviation in relation
to the alarm settings, the alarm can easily be displayed on a gas alarm unit. It is also possible to send information to the central operation control. The stabilizer can be delivered with either pressure transmitter
4–20 mA or with pressure switches. Gas reduction unit MM40 – STABILIZER must always be installed in compliance with the standards EN ISO 7396-1 and the appropriate national standards.
MM40 – STABILIZER is only a second stage reduction unit where the primary gas supply is provided by high pressure gas manifolds (such as MM40 – HP Unit). In case of a signal for pressure deviation in relation
to the alarm settings, the alarm can easily be displayed on a gas alarm unit. It is also possible to send information to the central operation control. The stabilizer can be delivered with either pressure transmitter
4–20 mA or with pressure switches. Gas reduction unit MM40 – STABILIZER must always be installed in compliance with the standards EN ISO 7396-1 and the appropriate national standards.
FOR A COMPLETE MM40 STABILIZER MANIFOLD ADD:
FOR A COMPLETE MM40 STABILIZER MANIFOLD ADD:
Gases | O2, Air, Air-800, N2, N2O, CO2 |
---|---|
Nominal flow (m3/h) | 40 |
Inlet maximal pressure | 16 (1600 kPa) |
Outlet nominal pressure | 4,5 bar (setting range 0,5–10 bar) |
Inlet connection 1 | G1/2"M + soldering piece pipe ø 12 mm |
Inlet connection 2 | Optional (G1/2"M + soldering piece pipe ø 12 mm) |
Outlet connection | G1/2"M + soldering piece pipe ø 12 mm |
Pressure relieve valve | Optional (6,8 bar; outlet pipe ø 15 mm) |
Pressure sensors | Optional (Pressure switches; Transmitter 4–20 mA) |
Emergency QC inlet | Optional QC by national standards |
Loading... ... loading table ... |
Item No. | Denomination | Gas | PRV | Signalization |
---|---|---|---|---|
0727329 | MM40 – Stabilizer | O2, N2O, Air, CO2, N2 | ||
0727337 | MM40 – Stabilizer | Air–800 | ||
0727332 | MM40 – Stabilizer | O2, N2O, Air, CO2, N2 | With pull-ring | Pressure switch |
0727344 | MM40 – Stabilizer | O2, N2O, Air, CO2, N2 | With pull-ring | 4-20mA |
0727343 | MM40 – Stabilizer | O2 | ||
0727346 | MM40 – Stabilizer | Air | ||
0727347 | MM40 – Stabilizer | Air–800 | ||
0727348 | MM40 – Stabilizer | CO2 | ||
0727364 | MM40 – Stabilizer | N2O |