The GCE MM90 Standby gas manifold is designed to be used as a third source of supply in medical central gas systems.
The GCE MM90 Standby manifold delivers gas when the nominal supply system pressure drops below 7 bar, serving as a backup source. The MM90 Standby should always be paired with the MM90 HP Unit and an alarm system to comply with standard requirements. It is recommended to use a stabilizer as the second stage.
The GCE Gas Alarm offers both visual and audible alerts, monitoring and signaling when the outlet pressure is too high, too low, or when a cylinder is empty, ensuring continuous safety and operation.
The GCE MM90 Standby manifold delivers gas when the nominal supply system pressure drops below 7 bar, serving as a backup source. The MM90 Standby should always be paired with the MM90 HP Unit and an alarm system to comply with standard requirements. It is recommended to use a stabilizer as the second stage.
The GCE Gas Alarm offers both visual and audible alerts, monitoring and signaling when the outlet pressure is too high, too low, or when a cylinder is empty, ensuring continuous safety and operation.
MM90 Standby includes the following components:
MM90 Standby backup gas manifold with pressure switches, purge valves, and HP filters
For a complete MM90 Standby manifold add:
MM90 Standby includes the following components:
MM90 Standby backup gas manifold with pressure switches, purge valves, and HP filters
For a complete MM90 Standby manifold add:
Gas type(s) | O2, Air, N2, N2O, CO2 |
---|---|
Nominal flow (m3/h) | 90 |
Inlet nominal pressure | 200 (20 000 kPa) |
Outlet nominal pressure | 7 bar (setting range 7–15 bar) |
Inlet connection | W21,8×1/14"M |
Outlet connection | G3/4"F + soldering piece pipe ø 22 mm |
Pressure relive valve | 16 bar |
Pressure relive valve pipe dimension | ø 10 mm |
Purge valves connection | W21,8×1/14"M + soldering piece pipe ø 10 mm |
Loading... ... loading table ... |
Item No. | Denomination | Gas | PRV | Signalization |
---|---|---|---|---|
0727307 | MM90 STANDBY | O2, Air | Standard | Pressure switches |
0727312 | MM90 STANDBY | N2O, CO2 | Standard | Pressure switches |
0727338 | MM90 STANDBY | N2O, CO2 | With pull-ring | Pressure switches |
0727367 | MM90 STANDBY | O2, Air | Standard | 4-20mA pressure transmitter |
0727368 | MM90 STANDBY | N2O, CO2 | Standard | 4-20mA pressure transmitter |